Silt Density Index Test
Instructions
Equipment
-
Silt Density Index Test Assembly (AMI DIRECT SDI - Model No. SDI-2000)
-
Filter Holder
-
Pressure Gauge
-
Pressure Regular
-
Ball Valve
-
Graduated Cylinder
-
Thermometer
-
Tubing
-
Filter Tweezers
-
Stop watch or other means of timing (not included)
Procedure
-
Unbolt the filter holder, wet it, and place a 0.45 micron
filter on the back-up plate using the dull tweezers.
-
Place the "O"-ring properly and then replace the top half of
the filter holder and bolt loosely.
-
Connect apparatus to the feedwater line (minimum 30 PSIG) in a
vertical, downflow position and open the needle valve a crack.
-
Loosen two adjacent filter-holder bolts, tilt the apparatus
and bleed out all trapped air, then retighten bolts and adjust the pressure to
30 PSI while starting the stopwatch.
-
Immediately run the flow into the graduated cylinder and
measure the time required to collect 500 ml, maintaining a pressure of 30 PSIG
during the run. Mark the time (t1)
and keep the watch running.
-
Repeat the previous step immediately after 5, 10, and/or 15
minutes of total elapsed time. Mark the collection times (t5,
t10,
tF).
-
Close the sample valve and disconnect the apparatus after the
15 minute reading. Remove the disk and place it in a plastic bag for later
examination.
Calculation
Calculate the ratios R = t1/t5
, t1/t10
and/or t1/tF.
Estimate % P30
at the selected time(s) from the following equation:
% P30
= [1-R] × 100.
Estimate the Silt Density Index, SDI, from the following equation:
SDI = % P30
÷ Elapsed Time (minutes)
The 15-minute index will generally be the lowest of the three, and
should be used for filter sizing purposes.
|